Tester Characteristics
◆ Simple operation, Chinese display, printing
◆ Convenient to use, no conversion needed
◆ The variable frequency dielectric loss tester has its own high voltage, strong anti-interference capability, and short test time advantages
Tester Features
| 1. | Due to the use of frequency conversion technology, accurate measurement can be ensured under strong electric field interference. |
| 2. | The instrument uses a Chinese menu operation, and the microcomputer automatically completes the entire measurement process. |
| 3、 | This instrument is also suitable for measuring the tgδ and capacitance of high-voltage electrical equipment in workshops, laboratories, and scientific research units; it has the advantages of convenience, simplicity, and accuracy for insulation oil loss testing. |
| 4、 | The instrument can measure high-voltage electrical equipment that is ungrounded or directly grounded using positive and reverse connection methods. It can also measure the tgδ of capacitive voltage transformers and the capacitances of main capacitors C1 and C2. |
| 5、 | The instrument is internally equipped with a high-voltage step-up transformer and incorporates safety protection measures such as zero-crossing closing and lightning protection. During the test, it outputs high voltages of different levels from 0.5KV to 10kV, making operation simple and safe. |
Tester Parameters
1、High voltage output: 0.5 ~10kV, each step increases by 500V, with a total of 20 steps
High voltage output cable, can be used on the ground
Three frequencies: 45Hz, 50Hz, 55Hz
2、Capacity: 1500VA
3、Accuracy: tgδ: ±(reading*1.0%+0.08%)
Cx: ±(reading*1.5%+5PF)
4、Resolution: tgδ:0.01% Cx:1pF
5、Measurement range: 0.1% < tgδ < 50%
3PF< Cx < 60000PF
At 10KV, Cx ≤30000PF
At 5KV, Cx ≤60000PF
6、Power supply: AC 220V±10% 50±1Hz
7、Harmonic adaptation: ≤3%
8、Operating conditions: -15℃-50℃, relative humidity <80%
9、Dimensions: 460(L)×335(W)×340(H)
10、Weight: 30 kg