Technical Parameters



1. Input voltage: AC 220V or 380V, 50Hz, automatic identification of input voltage.



2. Output voltage:



When input AC 220V, output 0~550V;



When input AC 380V, output 0~1000V.



Current: Maximum instantaneous current is 240A when measuring transformation ratio.



15A when testing volt-ampere characteristics.



3. Digital voltmeter measurement range: 0~999.9V; resolution: 0.1V.



When displaying current: 0~999.9mA; resolution: 0.1mA.



4. Digital ammeter measurement range: 0~15A (240A), resolution: 0.01A.



5. When measuring CT transformation ratio, the ratio can be automatically calculated, and the data and calculation results can be printed.



6. When testing volt-ampere characteristics, store and print 20 sets of current and voltage values and the volt-ampere characteristic curve.