Introduction to Anti-Interference Dielectric Loss Tester:
On-site anti-interference dielectric loss measurement, or precise dielectric loss measurement in the laboratory. The instrument is an integrated structure, with built-in dielectric loss bridge, variable frequency power supply, test transformer, and standard capacitor, etc. It adopts variable frequency anti-interference and Fourier transform digital filtering technology, fully automatic intelligent measurement, and the measurement data is very stable under strong interference. The measurement results are displayed on a large screen LCD, and a built-in micro printer can print out the results.
1.1 Main Technical Indicators
Accuracy: Cx: ±(reading×1%+1pF) TD2690E variable frequency anti-interference dielectric loss tester
tgδ: ±(reading×1%+0.00040)
Anti-interference index: variable frequency anti-interference, can still achieve the above accuracy under 200% interference
Capacitance range: Internal high voltage: 3pF~60000pF/10kV 60pF~1μF/0.5kV
External high voltage: 3pF~1.5μF/10kV 60pF~30μF/0.5kV
Resolution: 0.001pF, 4 significant digits
tgδ range: unlimited, resolution 0.001%, automatic identification of three types of test objects: capacitor, inductor, and resistor.
Test current range: 10μA~5A
Internal high voltage: set voltage range: 0.5~10kV
Output current: 200mA
Voltage regulation method: continuous smooth adjustment
Voltage accuracy: ±(1.5%×reading+10V)
Voltage resolution: 1V
Test frequency: 45, 50, 55 single frequency
Frequency accuracy: ±0.01Hz
External high voltage: test current 1A for normal wiring, power frequency or variable frequency 40-70Hz
Test current 10kV/1A for reverse wiring, power frequency or variable frequency 40-70Hz
CVT self-excitation method low voltage output: output voltage 3~50V, output current 3~30A
Measurement time: about 40s, depending on the measurement method
Input power: 180V~270VAC, 50Hz/60Hz±1%, mains or generator power supply
Computer interface: standard RS232 interface
Printer: Weihuang A7 thermal micro printer
Ambient temperature: -10℃~50℃