Working Principle



Under the action of AC voltage, the dielectric consumes a portion of electrical energy, which is converted into heat energy and causes loss. This energy loss is called dielectric loss. When an AC voltage is applied to the dielectric, there is a phase angle difference Ψ between the voltage and current in the dielectric. The complementary angle δ of Ψ is called the dielectric loss angle, and the tangent of δ, tgδ, is called the dielectric loss tangent. The tgδ value is a parameter used to measure dielectric loss. The instrument's measurement circuit includes a standard circuit (Cn) and a test circuit (Cx). The standard circuit consists of a built-in high-stability standard capacitor and the measurement circuit, while the test circuit consists of the test object and the measurement circuit. The measurement circuit is composed of a sampling resistor, a preamplifier, and an A/D converter. Through the measurement circuit, the amplitude and phase of the current in the standard circuit and the test circuit are respectively measured. Then, using a microcontroller with digital real-time acquisition methods and vector operations, the capacitance value and dielectric loss tangent of the test object can be obtained.



The instrument has built-in anti-interference measures to ensure accurate measurement under external electric field interference.



Technical Parameters











































































































































1



Operating Conditions



-5℃∽40℃



RH<80%



2



Anti-interference Principle



Variable Frequency Method



3



Power Supply



AC 220V±10%



Frequency Unlimited



4



High Voltage Output



0.5KV∽10KV



Every 0.1kV



Accuracy



2%


   

Maximum Current



200mA


   

Capacity



1500VA


   

5



Self-Excited Power Supply



AC 0V∽50V/15A



45HZ/55HZ



6



Resolution



tgδ: 0.001%



Cx: 0.1pF



7



Accuracy



△tgδ:±(Reading*1.0%+0.040%)



△C x :±(Reading*1.0%+1.0PF)


   

8



Measurement Range



tgδ



Unlimited



C x



15pF < Cx < 200nF


   
 

10KV



Cx < 60 nF


   
 

5KV



Cx < 100 nF


   
 

1KV



Cx < 200 nF


   

CVT Test



Cx < 200 nF


   

9



CVT Ratio Range



10∽10000



CVT Ratio Accuracy



0.1%


 

CVT Ratio Resolution



0.1


 

10



Dimensions



430(L)×330(W)×330(H)



11



Memory Size



80 Groups, Supports USB Data Storage



12



Weight



28 Kg