1. Instrument Measurement Range and Accuracy:
1) Capacitance Measurement
a. Capacitance Measurement Range: 0.1μF to 3,300μF
b. Reactive Power Measurement Range: 5 to 20,000 kvar
c. Measurement Accuracy: ±(1.0% reading + 0.02μF)
d. Resolution: 0.001μF
2) Current Measurement
a. Current Measurement Range: 0 to 199.9 mA; 0 to 1.999 A; 0 to 19.99 A; 0 to 199.9 A; 0 to 1000 A
b. Measurement Accuracy: ±(3.0% reading + 0.05 A)
c. Resolution: 0.01 A
3) Inductance Measurement
a. Inductance Measurement Range: Small inductance mode: 0.1 mH to 5 mH
Large inductance mode: 5 mH to 50 H
b. Inductive Reactance Measurement Range: 50 mΩ to 20 kΩ
c. Measurement Accuracy: ±(3.0% reading + 0.05 mH)
d. Resolution: 0.01 mH
4) Resistance Measurement Range:
a. Resistance Measurement Range: Small resistance mode: 50 mΩ to 1 Ω
Large resistance mode: 1 Ω to 20 kΩ
b. Measurement Accuracy: ±(3.0% reading + 0.05 Ω)
c. Resolution: 0.01 Ω
2. Power Supply
1) Rated Voltage: 220V AC ±10%
2) Rated Frequency: 50 Hz
3) Rated Output: 2V/20V/500VA
3. Operating Conditions:
1) Ambient Temperature: -10°C to +50°C
2) Relative Humidity: ≤90%
4. Printing Method: LCD display with full Chinese characters
Panel-mounted high-speed printer
5. Dimensions and Weight: 370×260×220 mm / 12 kg